Graphic Cases for Cannulated Screw Sets recall in Canada: incorrect measurement
In plain language
This recall involves Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets sold in Canada. They are being recalled because the screw gauges incorrectly measure screws as shorter than their actual length. This could lead to a surgeon inserting a screw that is shorter than intended.
What to do
- Stop using the product immediately.
- Check the model and lot numbers.
- Follow the instructions from DePuy Synthes for return or replacement.
Get alerts for recalls like this
Get email alerts when new recalls affect products in this category.
Affected products
Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets
Model: 690.375-US, 690.379-US, 690.380-US
Lot: FL00860, H340372
Do I have this product?
This recall only applies to specific products. Follow the steps below to check.
- 1
Check the product name
Make sure your product name matches one of the affected products listed above.
- 2
Check the lot number
Lot numbers identify specific production batches.
Affected lots: FL00860, H340372
- 3
Check the model number
Model numbers are typically on the product label or packaging.
Affected models: 690.375-US, 690.379-US, 690.380-US
If your product matches one or more of the details above, it may be affected by this recall.
If your product does not match these details, it is not affected by this recall.
When in doubt, always check the official notice.
Frequently Asked Questions
Why was this product recalled?
This recall involves Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets sold in Canada. They are being recalled because the screw gauges incorrectly measure screws as shorter than their actual length. This could lead to a surgeon inserting a screw that is shorter than intended.
What should consumers do?
Stop using the product immediately. Check the model and lot numbers. Follow the instructions from DePuy Synthes for return or replacement.
Where was the product sold?
This product was sold in Canada. For specific retailers or distribution areas, please check the official recall notice linked on this page.
Is there a health risk?
A potential safety risk has been identified with this product. Stop using it immediately and follow the instructions in the official recall notice.
When was this product recalled?
This product was recalled on March 23, 2018.
Is the DePuy Synthes brand affected by this recall?
Yes, DePuy Synthes products are affected by this recall. This recall involves Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets sold in Canada. They are being recalled because the screw gauges incorrectly measure screws as shorter than their actual length. This could lead to a surgeon inserting a screw that is shorter than intended.