RecallGuide.ca
    Medical devicesUpdated Jan 3, 2026QuickDrawOfficial source

    QuickDraw Venous Cannula recall in Canada: Risk of blood loss

    Share:

    In plain language

    This recall involves QuickDraw 22 and QuickDraw 25 Venous Cannulas sold in Canada. They are being recalled because using them for longer than 6 hours can cause the cannula to separate from its connector, leading to significant blood loss. Stop using the product and follow the return steps in the official notice.

    What to do

    • Stop using the product immediately.
    • Check model numbers QD22 and QD25.
    • Contact Edwards Lifesciences for further instructions.

    Get alerts for recalls like this

    Get email alerts when new recalls affect products in this category.

    Categories (optional)

    No spam. Unsubscribe anytime. We only email when there's a relevant recall.

    Affected products

    • QuickDraw 22 Venous Cannula

      Model: QD22

      Lot: All lots since Nov 14/17

    • QuickDraw 25 Venous Cannula

      Model: QD25

      Lot: All lots since Nov 14/17

    Why this matters

    Risk of significant blood loss

    Do I have this product?

    This recall only applies to specific products. Follow the steps below to check.

    1. 1

      Check the product name

      Make sure your product name matches one of the affected products listed above.

    2. 2

      Check the lot number

      Lot numbers identify specific production batches.

      Affected lots: All lots since Nov 14/17

    3. 3

      Check the model number

      Model numbers are typically on the product label or packaging.

      Affected models: QD22, QD25

    If your product matches one or more of the details above, it may be affected by this recall.

    If your product does not match these details, it is not affected by this recall.

    When in doubt, always check the official notice.

    Want alerts if similar products are recalled in Canada?

    Frequently Asked Questions

    Why was this product recalled?

    Risk of significant blood loss

    What should consumers do?

    Stop using the product immediately. Check model numbers QD22 and QD25. Contact Edwards Lifesciences for further instructions.

    Where was the product sold?

    This product was sold in Canada. For specific retailers or distribution areas, please check the official recall notice linked on this page.

    Is there a health risk?

    Yes, there may be a health risk. Risk of significant blood loss

    When was this product recalled?

    This product was recalled on January 10, 2020.

    Is the QuickDraw brand affected by this recall?

    Yes, QuickDraw products are affected by this recall. This recall involves QuickDraw 22 and QuickDraw 25 Venous Cannulas sold in Canada. They are being recalled because using them for longer than 6 hours can cause the cannula to separate from its connector, leading to significant blood loss. Stop using the product and follow the return steps in the official notice.