VITROS CHEMISTRY PRODUCTS (2020-12-21)
In plain language
This recall involves several VITROS CHEMISTRY PRODUCTS medical devices used for laboratory testing. They are being recalled because a medication, eltrombopag, can interfere with the test results, leading to inaccurate readings. Healthcare professionals should stop using the affected products and follow the return instructions in the official notice.
What to do
- Stop using the affected VITROS CHEMISTRY PRODUCTS immediately.
- Identify affected products by model or catalog number.
- Contact Ortho-Clinical Diagnostics Inc. for further instructions.
Get alerts for recalls like this
Get email alerts when new recalls affect products in this category.
Affected products
VITROS CHEMISTRY PRODUCT SYSTEM-TP SLIDES (TOTAL PROTEIN)
Model: 839 2292
Lot: All lots
VITROS CHEMISTRY PRODUCTS SYSTEM - BUBC SLIDES (BILIRUBIN)
Model: 161 2365, 838 3051
Lot: All lots
VITROS CHEMISTRY PRODUCTS SYSTEM – TBIL SLIDES (TOTAL BILIRUBIN)
Model: 815 9931
Lot: All lots
VITROS XT CHEMISTRY PRODUCTS-ALB TP SLIDES
Model: 684 4293
Lot: All lots
VITROS XT CHEMISTRY PRODUCTS TBIL-ALKP SLIDES
Model: 684 4296
Lot: All lots
Why this matters
Inaccurate test results due to drug interference
Do I have this product?
This recall only applies to specific products. Follow the steps below to check.
- 1
Check the product name
Make sure your product name matches one of the affected products listed above.
- 2
Check the lot number
Lot numbers identify specific production batches.
Affected lots: All lots
- 3
Check the model number
Model numbers are typically on the product label or packaging.
Affected models: 161 2365, 684 4293, 684 4296, 815 9931, 838 3051, 839 2292
If your product matches one or more of the details above, it may be affected by this recall.
If your product does not match these details, it is not affected by this recall.
When in doubt, always check the official notice.
Frequently Asked Questions
Why was this product recalled?
Inaccurate test results due to drug interference
What should consumers do?
Stop using the affected VITROS CHEMISTRY PRODUCTS immediately. Identify affected products by model or catalog number. Contact Ortho-Clinical Diagnostics Inc. for further instructions.
Where was the product sold?
This product was sold in Canada. For specific retailers or distribution areas, please check the official recall notice linked on this page.
Is there a health risk?
Yes, there may be a health risk. Inaccurate test results due to drug interference
When was this product recalled?
This product was recalled on January 15, 2021.
Is the VITROS CHEMISTRY PRODUCTS brand affected by this recall?
Yes, VITROS CHEMISTRY PRODUCTS products are affected by this recall. This recall involves several VITROS CHEMISTRY PRODUCTS medical devices used for laboratory testing. They are being recalled because a medication, eltrombopag, can interfere with the test results, leading to inaccurate readings. Healthcare professionals should stop using the affected products and follow the return instructions in the official notice.