RecallGuide.ca
    Medical devicesUpdated Jan 3, 2026SysmexOfficial source

    X-CAL, Sysmex E-Check, 8Check 3WP X-TRA Control, Sysmex SCS-1000 Universal Calibrator, XN CHECK, XN CAL, XN CAL PF, XN CAL - CUSTOMER SET, XN CAL PF - CUSTOMER SET, and XN-L CHECK (2018-05-08)

    Share:

    In plain language

    This recall involves several Sysmex hematology controls and calibrators sold in Canada. They are being recalled because they may provide inaccurate platelet counts if not mixed sufficiently. Healthcare professionals should stop using the products and follow the instructions in the official notice.

    What to do

    • Stop using the affected products immediately.
    • Check the lot or serial numbers of your products.
    • Remix the product and repeat analysis as instructed in the official notice.

    Get alerts for recalls like this

    Get email alerts when new recalls affect products in this category.

    Categories (optional)

    No spam. Unsubscribe anytime. We only email when there's a relevant recall.

    Affected products

    • X-CAL

      Model: 110-4008-0, 110-4009-0

      Lot: beginning with 7283

    • Sysmex E-Check

      Model: 199-4004-1, 199-5001-0, 199-5002-0, 201-6001-0, 201-6002-0, 201-6003-0, 201-6004-0

      Lot: beginning with 7269, beginning with 7277, beginning with 7305

    • 8Check 3WP X-TRA Control

      Model: 14030040, 14030060

      Lot: beginning with 7255

    • Sysmex SCS-1000 Universal Calibrator

      Model: 160-4002-0

      Lot: beginning with 7283

    • XN Check

      Model: 213499

      Lot: beginning with 7261

    • XN CAL

      Model: XN CAL

      Lot: beginning with 7282

    • XN CAL PF

      Model: XN CAL PF

      Lot: beginning with 7282

    • XN CAL - CUSTOMER SET

      Model: 213527

      Lot: beginning with 7283

    • XN CAL PF - CUSTOMER SET

      Model: 213536

      Lot: beginning with 7282

    • XN-L CHECK

      Model: 213569

      Lot: beginning 7300

    Why this matters

    Inaccurate test results

    Do I have this product?

    This recall only applies to specific products. Follow the steps below to check.

    1. 1

      Check the product name

      Make sure your product name matches one of the affected products listed above.

    2. 2

      Check the lot number

      Lot numbers identify specific production batches.

      Affected lots: beginning 7300, beginning with 7255, beginning with 7261, beginning with 7269, beginning with 7277, beginning with 7305, beginning with 7282, beginning with 7283

    3. 3

      Check the model number

      Model numbers are typically on the product label or packaging.

      Affected models: 110-4008-0, 110-4009-0, 14030040, 14030060, 160-4002-0, 199-4004-1, 199-5001-0, 199-5002-0, 201-6001-0, 201-6002-0, 201-6003-0, 201-6004-0, 213499, 213527, 213536, 213569, XN CAL, XN CAL PF

    If your product matches one or more of the details above, it may be affected by this recall.

    If your product does not match these details, it is not affected by this recall.

    When in doubt, always check the official notice.

    Want alerts if similar products are recalled in Canada?

    Frequently Asked Questions

    Why was this product recalled?

    Inaccurate test results

    What should consumers do?

    Stop using the affected products immediately. Check the lot or serial numbers of your products. Remix the product and repeat analysis as instructed in the official notice.

    Where was the product sold?

    This product was sold in Canada. For specific retailers or distribution areas, please check the official recall notice linked on this page.

    Is there a health risk?

    Yes, there may be a health risk. Inaccurate test results

    When was this product recalled?

    This product was recalled on August 17, 2018.

    Is the Sysmex brand affected by this recall?

    Yes, Sysmex products are affected by this recall. This recall involves several Sysmex hematology controls and calibrators sold in Canada. They are being recalled because they may provide inaccurate platelet counts if not mixed sufficiently. Healthcare professionals should stop using the products and follow the instructions in the official notice.